Precision PCx Test Strips
The following information regarding the Precision PCx and Precision
PCx Plus Test Strip can be found in the test strip instructions for use:
System Operating Ranges
Storage and Handling
Quality Control
Limitations of the Procedure
Performance Characteristics
Precision PCx Monitor
Dimensions:
Length: 19.7 cm (7.7 in)
Width: 7.5 cm (2.95 in)
Thickness: 5.1 cm (2.0 in)
Weight: 280 grams (10 ounces)
Power Source: Two standard alkaline AA batteries or
optional MediSense rechargeable battery
pack.
Battery Life:
Battery performance is a combination function of how
often the meter is used and the duration time that testing
is conducted.
Alkaline Battery:
(2) cells 1.5 Volts, 2.4 Amp Hours typical
capacity
Rechargeable Battery Pack:
(2) cells 1.2 Volts, 1.1 Amp Hours
minimum capacity
Note: The MediSense Rechargeable Battery Pack can
be recharged for 500 full charge cycles. A
considerably higher battery cycle life can be
obtained if the Precision PCx monitor is placed in
the docking station at the end of each use.
Memory:
Patient Test 4000
Control Test 1000
Proficiency Test 60
Linearity Test 20 (5 levels, 4 replicates per level)