8-26
What You See
on the Display
What You Do Comments
4. The administrator will
select one of the
following options:
Press 1 – Allow Test.
Press 2 – Warn.
Press 3 – Lockout.
The display shows
the QC Interval Security
menu. The operator can
choose between QC
intervals. The settings
are:
• Time Interval – QC
Interval based on a set
number of hours
between control tests.
• Time of Day – choose
up to three times
during the day that QC
Interval will take place.
• Number of Patient
Tests – QC Interval
based on the number
of patient tests
specified.
5. The administrator will
select one of the
following options:
Press 1 – Time Interval:
continue to step 6.
Press 2 – Time of Day:
skip to step 7.
Press 3 – Patient Tests:
skip to step 8.
QC Time Interval option,
the Time of Day option,
or the Number of Patient
Tests option.
6. Enter a Time Interval
number between 1-99,
then press Enter. Skip
forward to step 9.
The display returns to
the Security QC Interval
menu.
7. Enter up to 3 times
of the day in
chronological order,
starting at midnight.
Press Enter to confirm
each time. Skip forward
to step 9.
The display returns to
the Security QC Interval
menu.