This section describes the steps to be taken to test Control Solutions in
order to verify the performance of the Precision PCx Monitor and
Precision PCx and Precision PCx Plus Test Strips.
The quality control features of the monitor can be customized to fit the
requirements of each facility. The monitor can be set to require:
Tests of Low, Normal, and High Control Solutions or any variation of
this depending on the facility’s policy
Tests of Control Solutions at regular times or time intervals, or by
number of patient tests
Control test results will appear as Numeric or Pass/Fail.
These quality control features can be set through the Menu Mode menu
(Change Setup/Test Setup Menu).
Operating Guidelines for Control Solution Testing
Observe the following guidelines to obtain optimal quality control results
using the Precision PCx System:
Use only control solutions approved for use on the test strip instructions
for use to verify the performance of the Precision PCx and Precision
PCx Plus Test Strips and the Precision PCx Monitor.
When opening a new bottle, write the current date on the bottle label.
Each bottle of control solution is stable for 90 days after opening
(replace cap tightly after each use). Discard all unused solutions 90 days
after opening.
Invert the control solution bottle several times to ensure thorough
mixing before use.
Invert and tap the capped control solution bottle to remove air bubbles
from the nozzle of the bottle.
Do not scan one test strip packet’s barcode and use a test strip from
another packet. This may cause incorrect assay results to be generated.
Replace the correct cap on the bottle and tighten the cap immediately
after each use.
Store the control solutions at temperatures between 39º and 86ºF
(4º and 30ºC).
Do not use control solutions after the expiration date printed on the
bottles and the box. The Precision PCx Monitor does not accept control
solutions that have passed their expiration date.
When To Test Control Solutions
As required by your facility’s quality policy or local regulatory
requirements.
When you question blood glucose results
When you open a new lot of test strips
Control Test 4