Review/Change Test Setup for All Tests:
The options that are associated with the Review/Change Test Setup
for All Tests apply to all test types.
Use this menu to review or change these options for all tests.
Upload Interval
The monitor can be set up to warn or lockout a task if the monitor has
not been uploaded at a regularly scheduled interval. The upload interval
options are:
Allow Test–allows the testing to continue. No message will appear
on the monitor.
Warn–allows the testing to continue but the operator will receive a
message that the upload interval (as specified below) has expired.
Lockout–does not allow any testing until the data has been uploaded.
This option is set to designate a required maximum number of hours
or days (from 1 to 99) between data uploads. When a task is selected
and the upload interval has expired, the operator will receive a
message that the Upload Interval has expired. If Upload Interval is
expired, the monitor will prevent an operator from doing other tasks
until the stored data is uploaded.
Operator ID
The Operator ID menu has the following options which can be viewed
or changed:
Disabled, Optional or Required entry of an Operator ID for all testing.
Require a minimum and a maximum number of digits for the length
of an Operator ID.
Allow any combination of five common barcode formats used in
the facility:
I2 of 5 (with or without check digit)
Code 39 (with or without check digit, alphanumeric or
ASCII format)
Code 128
Codabar
Code 93
Truncate (ignore) leading and/or trailing barcode digits.
Note: The display shows selected options in bold type.