Questions? Refer to Page 10-13 For Customer Care Information.
8-55
What You See
on the Display
What You Do Comments
Add a Test Strip Lot to the Strip Lot List:
1. Press 1 Add. The display shows the
Scan or Enter Strip Code
screen.
2. Press Scan or manually
enter* the test strip lot
number via the keypad
and press Enter.
The display shows the
Strip Code option.
The operator can return to
the Strip Lot menu to add
additional test strip lots or
can continue to the next
menu to set minimum
and maximum control
solution ranges for the test
strip lot.
3. The administrator will
select one of the
following options:
Press 1 – Continue.
Press 2 – Change, to
change the control ranges.
The display prompts the
operator to enter in the
test strip lot minimum and
maximum values for the
low control test.
4. Key in the minimum
and maximum values
for the low control,
press Enter after each
value entered.
The display prompts the
operator to enter in the
test strip lot minimum and
maximum values for the
normal control test.
5. Key in the minimum
and maximum values
for the normal control,
press Enter after each
value entered.
The display prompts the
operator to enter in the
test strip lot minimum and
maximum values for the
high control test.
* Manual entry of strip lots with alpha-numeric barcodes is not supported.