What You See
on the Display
What You Do Comments
Questions? Refer to Page 10-13 For Customer Support Information.
3-5
4. Press Scan to scan the
Patient ID barcode, or
manually enter the
Patient ID via the
keypad, then press
Enter.
After entering the Patient
ID, the test strip lot screen
appears next.
5. Press Scan to scan the
test strip barcode, or
manually enter* the test
strip lot number via the
keypad, then press
Enter.
After the test strip lot
number has been entered,
the Sample Type screen
will appear if this setting
has been enabled. If it
does not appear, skip
down to step 7.
6. The operator can select
one of the following:
Press 1 – Capillary or
Arterial.
Press 2 – Venous.
Note: Use VENOUS
MODE only for venous
samples that have been in
capped tubes and have not
been mixed with air. Use
CAPILLARY or
ARTERIAL MODE for
all other samples.
7. Open the foil test strip
packet at the notch and
tear up or down to
remove the test strip.
* Manual entry of strip lots with alpha-numeric barcodes is not supported.