Precision PCx Plus Test Strip
8-32
What You See
on the Display
What You Do Comments
10. The administrator will
select one of the
following options:
Press 1 – Next, to return
to Test Setup Test Type.
Press 2 – Change, to
change action range.
If the action range is
disabled, the display is
returned to the Test Setup
Test Type screen. If the
action range is enabled,
the display shows the
action range.
The display shows the
Glucose Action Range
options.
11. The administrator will
select one of the
following options:
Press 1 – Disabled. Skip
forward to step 13.
Press 2 – Enabled.
The display shows the
Action Range screen.
When the action range
is enabled, the display
shows an Action Range
Low and High values
screen. Enter a low action
range value and a high
action range value.
13. Press 1 – Next to
return to the Test
Setup Test Type
Screen.
12. Enter the low value
and press Enter.
Enter the high value
and press Enter.
The display is returned to
the Glucose Action Range
screen.
Refer to test strip
instructions for use for
Action Range.
(1.1 mmol/L,
-33.3 mmol/L)
Precision PCx Test Strip